浩瀚色譜(山東)應用技術開發有限公司采用頂空-氣相色譜-質譜聯用技術,實現對乳及乳制品(液TAI乳、發酵乳、嬰幼兒配方乳粉、冰淇淋)中環氧乙烷和2-LV乙醇的準確定性和定量。樣品于頂空瓶中經稀釋定容后,90℃下平衡20 min,采用HH-624毛細管色譜柱(60 m×0.32 mm,1.8μm)程序升溫模式實現目標物色譜分離,質譜采用選擇離子掃描模式對環氧乙烷和2-LV乙醇進行檢測,基質匹配標準曲線法定量。結果表明:該方法在0.1~10.0 mg/L線性良好(R2≥0.99),環氧乙烷和2-LV乙醇在YE態乳、發酵乳和冰淇淋中的檢出限為0.12、0.60 mg/kg;在低、中、高3個加標水平下,2種化合物的平均回收率為75.9%~101.6%,相對標準偏差均小于10%。該方法精密度和準確度良好,且方法靈敏度高,滿足乳及乳制品復雜基質中環氧乙烷和2-LV乙醇的檢測要求。



乳及乳制品中環氧乙烷及2-LV乙醇的檢測毛細管柱HH-624  測試譜圖:



Determination of ethylene oxide and 2-chloroethanol in milk and dairy products by capillary column HH-624




Determination of ethylene oxide and 2-chloroethanol in milk and dairy products by capillary column HH-624 Details:






Name: Capillary Column


Stationary phase: 6% nitrile bing phenyl 94% dimethyl polysiloxane,


Specification: 60m * 0.32mm * 1.4um


Model: HH-624








Haohan Chromatography (Shandong) Application Technology Development Co., Ltd. uses headspace gas chromatography mass spectrometry technology to achieve accurate qualitative and quantitative determination of cyclohexane and 2-chloroethanol in milk and dairy products (liquid milk, fermented milk, infant formula milk powder, ice cream). After the sample is diluted and brought to volume in a headspace bottle, it is equilibrated at 90 ° C for 20 minutes, and a HH-624 capillary chromatographic column (60 m × 0.32 mm,1.8 μ m) The temperature programmed mode achieves chromatographic separation of the target substance. The mass spectrometry uses a selective ion scanning mode to detect ethylene oxide and 2-chloroethanol, and the matrix matching standard curve method is used for quantification. The results showed that the linearity of the method was good between 0.1 and 10.0 mg/L (R2 ≥ 0.99), and the detection limits of ethylene oxide and 2-chloroethanol in liquid milk, fermented milk, and ice cream were 0.12 and 0.60 mg/kg; At three spiking levels of low, medium, and high, the average recoveries of the two compounds were 75.9% to 101.6%, with relative standard deviations less than 10%. The method has good precision and accuracy, and high sensitivity, meeting the detection requirements for ethylene oxide and 2-chloroethanol in complex matrices of milk and dairy products.








Determination of ethylene oxide and 2-chloroethanol in milk and dairy products by capillary column HH-624 Test spectrogram: